limiting oxygen index apparatus Introduction To automatically determine the minimum quantity of oxyg
Электронная почта: info@standard-groups.com
oxygen indexer testing machine Technical Parameters:
Oxygen concentration measurement range: 0-90%.
Oxygen concentration accuracy: ± 0.1%
Flow meter range: 1.0 ~ 10.0L / min
Oxygen concentration measurement accuracy: ± 0.4%.
Response time: <10s
Combustion cylinder air flow:40mm/s ± 2mm/s; concentration adjustment accuracy: ± 0.1%.
Top air flow of the combustion cylinder: 90mm±10mm/s。
Portable igniter, downward jet of 16±4mm, adjustable.
Measurement time up to 5mins, accuracy of ± 0.5S.
Dual flow meter and pressure gauge device: accuracy of ± 1%.
Test temperature range: 25 ℃ ~ 150 ℃。
oxygen index testing instrument Features:
Imported oxygen concentration sensor (paramagnetic oxygen concentration sensor for model F248A, electrochemical oxygen concentration sensor for model F248D ) showing the oxygen concentration accuracy of ± 0.1%.
High temperature resistant quartz glass tube, outlet diameter 40mm.
Supportive and non-supportive sample holder.
Portable ignition, easy to operate.
Patented gas flow control unit which equipped with mass flow controller mixing nitrogen and oxygen can precisely control the gas flow.
After setting the oxygen concentration on screen, PLC control system can automatically adjusts the oxygen concentration according to the set value , precise and easy control.
The gas mixing chamber with patented design can ensure that the oxygen concentration measurement more accurate and the mixture of oxygen and nitrogen more better.
LCD real-time can display flow, time , test data and other parameters.
Dual flow meter and pressure gauge device: accuracy of ± 1%.
Equipped with special ignition.
Вышеуказанный контент предназначен для ссылки и общения только между инсайдерами отрасли и не гарантирует его точности или полноты. Согласно соответствующим законам и нормативным актам и правилам этого веб-сайта, подразделения или лица, приобретающие связанные с ними предметы, должны получить действительную квалификацию и условия квалификации.
Телефон компании
+86-21-6420 0566
Время работы
С понедельника по пятницу
Мобильный телефон:
13816217984
Электронная почта:
info@qinsun-lab.com