Closed - Space Tester for Aerosol Sprays - Qinsun
Специализированные производители контрольно - измерительных приборов
Национальная & quot; горячая линия & quot;:+86 021-64208466
Продукты

Closed - Space Tester for Aerosol Sprays

Tests flammability of aerosols when sprayed into an enclosed space.

WhatsApp: +86 13816217984

Email: info@qinsun-lab.com

Содержание

Описание продукции

Conforms to UN Recommendations on the Transportation of Dangerous Goods - Manual of Tests and Criteria 31.5 Confined Space Ignition Test and GB/T 21631-2008 Test Method for Ignition of Dangerous Goods Aerosols in a Confined Space.


Особенности продукции

1. It adopts an embedded processor and the Windows CE operating system. 2. With an 8 - inch liquid - crystal touch screen, it can display the test status in real - time. 3. It integrates two working modes: the "standard mode" and the "custom mode". 4. The aerosol can sprays automatically, and the aerosol can clamping device can automatically turn the can upright and upside - down. 5. Automatic timing during spraying. 6. With a dual - detection module consisting of a pressure sensor and a nanosecond - level photosensitive sensor, it can automatically detect whether the flame is burning or whether an explosion occurs in an enclosed space. 7. It is equipped with a ventilation system to promptly discharge the smoke and unburned aerosol in the drum, ensuring the accuracy of the test and the safety of the experimenters. 8. The test process is fully automated, requiring no manual intervention. It supports functions such as parameter setting, data storage, and report viewing. 9. It is equipped with a temperature and humidity module to monitor the temperature and humidity of the test environment in real - time.


Технические параметры

Operating Environment

(540)℃, <85%RH

Ambient Temperature Measurement Range

-10℃+60℃

Ambient Temperature Measurement Accuracy

±0.5℃

Ambient Humidity Measurement Range

5%RH95%RH

Ambient Humidity Measurement Accuracy

±5.0%RH

Combustion Detection Response Time

≤6ms

Pressure Measurement (Differential Pressure) Range

-5 kPa ~ 5kPa

Pressure Measurement Accuracy

±0.1%FS

Timing Display Resolution

0.1 сек.

Ignition Flame

Compliance with the standard candle flame, adjust the height of the cushion column 20mm / a

Cylindrical Vessel Dimensions

Ф600mm × 720mm

Positioning of Aerosol Tanks

Aerosol canisters can be automatically turned over

Dimensions of Compatible Misting Tanks

Height (120 ~ 400) mm, diameter (30 ~ 80) mm

Distance of the Cylindrical Vessel from the Nozzles of Misting Tanks

(25 ~ 100) mm adjustable

Liquid Accumulation Tray

accumulation of spray cans sprayed the remnants of the liquid can be withdrawn at any time to do the treatment


5603

Вышеуказанный контент предназначен для ссылки и общения только между инсайдерами отрасли и не гарантирует его точности или полноты. Согласно соответствующим законам и нормативным актам и правилам этого веб-сайта, подразделения или лица, приобретающие связанные с ними предметы, должны получить действительную квалификацию и условия квалификации.


Онлайн сообщение

Контактные данные

Телефон компании

+86-21-6420 0566

Время работы

С понедельника по пятницу

Мобильный телефон:

13816217984

Электронная почта:

info@qinsun-lab.com

QR-код